MIRROR-ANALYSER SF6 For verification of several parameters with only one measurement
Description
3-035R-R…
MIRROR-ANALYSER SF6
This multi-functional device allows the determination of up to three quality parameters with only one measurement:
- Moisture concentration (frost point / dew point)
- SF6 volume percentage
- SO2 concentration
The Mirror-Analyser SF6 uses the physical dew point mirror measuring principle which is characterised by its high precision and utmost reliability in order to measure the moisture content. By cooling the integrated mirror the moisture content of the gas is determined by measuring the temperature depending on the condensation or icing of the mirror. The other parameters of SF6 quality are determined by measuring the speed of sound (SF6 volume percentage) and electrochemical reaction (SO2 concentration).
- High accuracy and reliability in moisture determination (dew point mirror measuring principle)
- Results of up to 500 measurements can be stored with name, date and time
- No emissions of measuring gas (integrated gas return system)
- USB- and LAN connection
- Modular interchangeability of the sensors
- Adjustable user languages: DE, EN, FR, ES
- Low maintenance due to self-test functions
- Compact design, easy handling and transportation (installed in a trolley case)
- Easy and user-friendly menu navigation via high quality 7“ capacitive colour touch screen